• BACKGROUND
    • Metal-on-metal (MoM) revisions have shown high rates of complications from aseptic loosening, deep infection, and dislocation. The purpose of this study is to report on outcomes and complications of a large consecutive series of patients who were revised for failed MoM total hip arthroplasty (THA).
  • METHODS
    • We evaluated 188 patients (203 hips) who underwent revisions of failed MoM THA. Mean age at the revision was 60 years old. Gender was female in 112 hips (55%) and male in 91 hips (45%). Mean interval to failure after primary THA was 4.9 years (range, 0-18 years). The acetabular component was revised in 183 cases (92%). Clinical outcomes assessed included pain score and Harris hip score.
  • RESULTS
    • Mean follow-up from revision was 4.2 years. Harris hip score improved from 53.6 pre-revision to 73.5 at most recent follow-up (P < .001). Pain level significantly improved from 17.2 pre-revision to 32.8 post-revision (P < .001). Reoperations occurred in 28 (14%) hips. Re-revision was required in 16 hips (7.9%). The most common reasons for re-revision were aseptic loosening (5), dislocation (3), infection (2), and iliopsoas tendonitis (2). Re-revision was significantly higher in cases of pseudotumor. There was no difference in survival with ultraporous cups. Three hips required re-revision to custom triflange components. All-cause survival was 90.5% at 4.2 years. Metal ion levels significantly declined after revision.
  • CONCLUSION
    • Revisions of failed MoM THA showed improvements in clinical outcomes, but present significant surgical challenges and complications especially in cases with soft tissue damage and pseudotumor.