• OBJECTIVE
    • We hypothesize that cut screws will deform in a manner that increases the core and outer diameters of the screw hole compared to uncut controls, and effects will be more pronounced in titanium screws.
  • MATERIALS AND METHODS
    • We used biomechanical polyurethane foam blocks to simulate cortical bone. We organized four groups of stainless steel and titanium cut and uncut screws. Blocks were fitted with a jig to ensure perpendicular screw insertion. We imaged the blocks using digital mammography and measured them using PACS software. Power analysis determined a power of 0.95 and an alpha error of 0.05.
  • RESULTS
    • Highly statistically significant differences in core diameter were found after cutting stainless steel and titanium screws. Cutting stainless steel screws increased core diameter by 0.30 mm (95% CI, 0.16 to 0.45; p < .001). Titanium screws' core diameter increased by 0.45 mm (95% CI, 0.30 to 0.61; p < .001). No significant differences were found in the outer diameters of stainless steel and titanium screws after cutting.
  • CONCLUSION
    • Titanium and stainless steel screw tracts demonstrated screw core diameter and screw thread pattern deformation after cutting. Titanium screws demonstrated more significant effects.