• BACKGROUND
    • The purpose of this study is to assess the radiographic results, clinical outcomes, and survivorship of a modern cementless dual-mobility cup (DMC) implant as a primary THA with a minimum of 10 years follow-up.
  • METHODS
    • This study retrospectively assessed a series of 310 primary THAs using a modern-generation cementless DMC (Saturne acetabular cup) between April 2001 and December 2005 at a single center. Patients were followed prospectively clinically and radiographically after surgery. Hips with follow-up less than 120 months were excluded from the study (5 lost to follow-up and 167 deceased). In total, a cohort of 138 hips were included for preoperative and postoperative analysis with an average follow-up of 152.4 months. All complications were collected, and a Kaplan-Meier survival analysis was performed.
  • RESULTS
    • There was a significant increase in the mean Harris and Postel-Merle d'Aubigne scores between preoperative and postoperative cohorts (P < .001). No loosening of the cup and no acetabular osteolysis were found at final follow-up. No prosthetic dislocation, no intraprosthetic dislocation, and no infections were reported. The survival curve of THA in the total cohort (N = 310) was about 98% at 10 years with 3 stem revisions for femoral fracture. One psoas impingement was also described.
  • CONCLUSION
    • This study showed no acetabular component failure and no reported cases of acetabular osteolysis with this DMC acetabular component retention at 10 years. No prosthetic or intraprosthetic dislocation was reported.