• BACKGROUND
    • Conservative treatment of lateral clavicle fractures (Neer type II) often ends with unsatisfactory results for patients. Operative treatment is thus gaining acceptance. A retrospective study found success using a double plate technique for these injuries. For the current study, a standardized testing setup was developed to compare the mechanical properties of single versus double plate technique for lateral clavicle fractures.
  • METHODS
    • Six synthetic bones were tested for each technique. Neer type IIB fractures were created using computer-aided design (CAD). Fatigue testing was carried out with a cantilever bending test. Parameters measured were cycles undergone, failure load and stiffness at the point of failure.
  • RESULTS
    • The standardized testing setup was able to provide reproducible failures. The double plate technique lasted about 16,000 more cycles and withstood an additional 22.4 N on average.
  • CONCLUSIONS
    • The new modified standardized testing setup produced reproducible fixation failures for both clavicle fracture fixation techniques. The double plate technique seems to be mechanically superior to the single plate technique.