HIP-G Proximal Femur

Images
https://upload.orthobullets.com/topic/423332/images/64cce1b1-658b-4f30-a419-8caaf32a5ebb_screenshot_2026-02-23_at_11.37.49_am.jpg
https://upload.orthobullets.com/topic/423332/images/6e31e1cd-b4ab-4aad-9c2d-f57370cc3a56_hip_g_3.jpg
https://upload.orthobullets.com/topic/423332/images/204fe888-513c-49b9-b09e-d33ef130c244_hip-g.jpg
  • Summary
    • The HIP-G Proximal Femur Nail is an intramedullary nail system for treatment of fractures, revisions, and tumor stabilizations in the proximal femur, including combined injury patterns affecting the shaft area (long nails)
    • Key system concepts include a two-portal proximal fixation concept for rotational stability intra- and post-operatively, and a DyCon® screw pair intended to function as a dynamically controlled force carrier system in the femoral neck/head
    • Controlled dynamization is provided via the inferior DyCon® Set Screw engaging the groove of the Lag Screw, allowing scalable lateral dynamization (0mm static to 10mm dynamic+) while inhibiting medial migration via full contact area between screws
  • Indications
    • All nails
      • stable and unstable pertrochanteric fractures
      • intertrochanteric fractures
    • Additionally for long nails
      • subtrochanteric fractures
      • pathological fractures 
      • nonunions and malunions
    • Contraindications (general, as described)
      • special circumstances such as infection or marked local inflammation
      • conditions requiring physician risk/benefit consideration, including
        • dependence on patient bone quality for adequate fixation until bony healing occurs
        • compromised vascularity
        • excessive stresses on bone/implant (e.g., severe obesity) or degenerative diseases
        • interfering systemic conditions such as infection
        • demonstrated allergy or foreign body sensitivity to any implant materials
  • Anatomy
    • Osteology
      • proximal femur landmarks relevant to trochanteric entry nailing include
        • the greater trochanter
        • femoral neck/head
        • calcar region
        • proximal diaphysis
      • inferior femoral neck cortex (Adam’s arch/calar region) is a critical zone when planning inferior trajectories near the neck cortex
    • Muscles
      • abductors (gluteus medius/minimus) attach to the greater trochanter and influence trochanteric tenderness and lateral approach dissection planes
      • iliopsoas insertion on the lesser trochanter can contribute to flexion/external rotation of proximal fragments in subtrochanteric patterns
      • short external rotators and gluteal musculature contribute to deforming forces and can complicate reduction in unstable patterns
    • Ligaments
      • hip capsule and iliofemoral/pubo-femoral/ischiofemoral ligaments contribute to hip stability and can limit reduction maneuvers
      • ligamentum teres contributes minimally to adult femoral head perfusion compared with retinacular vessels
    • Nerves
      • sciatic nerve is at risk with excessive traction, aggressive posterior retraction, or malpositioning
      • superior gluteal nerve is relevant to proximal lateral dissection planes and abductor function
      • lateral femoral cutaneous nerve may be irritated with proximal lateral incisions and retraction
    • Blood supply
      • medial femoral circumflex artery (deep branch) supplies most of the femoral head via retinacular vessels
      • intracapsular displacement and excessive manipulation can threaten perfusion in femoral neck–adjacent injuries
  • Preoperative Planning
    • Imaging
      • obtain pelvic overview imaging and longitudinal imaging of the affected femur with the hip joint in a/p and (if feasible) lateral projections
  • Approach
    • Position patient supine on a table that allows unobstructed fluoroscopy and a freely adjustable extension device via foot fixation
      • position the healthy leg rotated to the opposite side with ~90° hip/knee flexion and supported
      • secure the arm on the affected side across the body
      • alternative
        • lateral position is possible (e.g., shaft fractures), acknowledging potentially more difficult proximal reduction
    • Localize incision using fluoroscopic marking for trochanter access
      • in a/p, place a drill-wire transverse to the body axis resting on the trochanter tip in projection
      • in lateral, place drill-wire centrally in femoral axis and mark longitudinally above the horizontal line
    • Perform a proximal lateral incision for trochanter entry
      • create ~2–3cm incision starting ~2–4cm proximal to the trochanteric apex (soft tissue dependent)
      • sharply split fascia longitudinally; elevate musculature to palpate the anterior-to-posterior trochanteric ridge
      • divide the ridge into thirds; usual entry point is at the junction of the anterior third and middle third
  • Technique
    • Reduction & canal opening
      • reduction
        • achieve closed reduction (length, axis, rotation) under fluoroscopy
        • open reduction is required if displaced subtrochanteric patterns are irreducible.
      • entry point
        • place drill-wire centrally in the femoral axis (check valgus path in A/P)
        • correction
          • use parallel holes in the conical sleeve if repositioning is needed
      • opening
        • advance tissue protection sleeve to the trochanter
        • ream over the drill-wire until the depth stop hits the sleeve
          • if the canal is narrow, use a ball-tip guide wire and ream over it
    • Nail insertion & setup
      • selection
        • match nail to length, diameter, and CCD angle (120°/125°/130°)
      • insertion
        • attach nail to targeting arm and double-check the holding bolt is fully tightened
        • advance nail with slight rotation
      • sleeves
        • insert K-wire sleeves through the master sleeve until they firmly contact the lateral cortex
        • secure via the ratchet mechanism
    • Two-portal drill-wire placement
      • superior (lag screw)
        • center the femoral head in lateral view
        • drill wire to a central position
        • final tip should be ~5mm below the bone-cartilage transition in A/P view
      • inferior (DyCon set screw)
        • place the inferior wire parallel to the first
          • anchor it 5-10mm shorter than the superior wire
    • Lag screw insertion
      • measurement
        • measure length via ruler
        • drilling depth = measured value minus 5mm
      • drilling
        • remove superior K-wire sleeve (keep inferior wire in for stability)
        • drill over the superior wire until the stop is reached
      • insertion
        • insert cannulated lag screw
        • align markings on the driver with the sleeve
        • final positioning is adjustable in 1.6mm increments
        • remove driver and superior wire
    • Set screw fixation (DyCon or Axial)
      • DyCon (inferior)
        • replace inferior K-wire sleeve with the set screw sleeve
        • drill until the stop
        • insert DyCon screw
          • it will automatically "push off" the driver at the 5mm dynamization depth
      • Axial (top-down)
        • insert via the nail holding bolt pathway
        • tighten until it engages the lag screw groove
          • do not back out
    • Distal locking
      • short/intermediate (guided)
        • use the targeting module
        • drill with Ø4.2mm bit through the sleeve (ratchet seated on bone)
        • add length for the medial cortex and insert Ø5mm screw
      • long nails (freehand)
        • obtain "perfect circles" on fluoroscopy
        • static
          • use two round holes
        • dynamic 
          • use the lower position of the elongated hole
    • Completion & optional steps
      • disassembly
        • remove sleeves and unscrew the nail holding bolt only after all locking is complete
      • closure
        • verify final implant position and screw lengths via fluoroscopy before wound closure
      • end caps (optional)
        • used only if nail extension or scar tissue protection is required
  • Technical specifications
    • Nail lengths 
      • short
        • 180mm
      • intermediate
        • 240mm
      • long
        • 260–480mm (multiple lengths)
    • Nail diameter
      • short distal diameter
        • 10mm
      • intermediate distal diameters
        • 10mm 
        • 12mm
      • long distal diameters
        • 9mm
        • 10mm
        • 12mm
        • 14mm
    • CCD angles available
      • 120°
      • 125°
      • 130°
    • Anteversion
      • 12° anteversion of femoral neck (as described for system concept/rotation control)
    • Proximal geometry/features include
      • proximal diameter Ø15.7mm
      • 4° lateral bend
      • two proximal portals for lag screw (superior) and set screw (inferior) trajectories
      • self-retaining mechanism for DyCon® Set Screw (as described)
    • Antecurvature 
      • short and intermediate nails use average femoral antecurvature of 1100mm
      • long nails have length-dependent antecurvatures
    • distal locking options include
      • Ø5mm screws
      • horizontally oriented true M-L configuration
      • static and dynamic locking options 
  • Screws
    • Lag screw
      • diameter Ø10.5mm
      • lengths 70–130mm (5mm increments)
    • Dycon® set screw (inferior set screw of dycon® screw pair)
      • controlled lateral dynamization control (0–10mm) while inhibiting medial migration via screw-to-screw contact
      • 1:1 relationship with lag screw
      • short dycon® set screw
        • can be used universally when needed/preferred
    • Axial set screw
      •  can be used alone or with dycon® set screw
        • if used alone, medial movement of lag screw is blocked and lateral movement is free depending on remaining groove length of lag screw
    • Distal locking screws
      • diameter
        • Ø5.0mm
      • lengths
        • 25–120mm
    • End cap screws
      • 0mm
      • 5mm
      • 10mm
      • 15mm
      • 20mm