Tibial Shaft Fractures
Updated: Feb 26 2025

Pangea™ Distal Medial Tibia Plate

Images
https://upload.orthobullets.com/topic/423209/images/81cbc3fb-75bf-48d1-8a1c-b3b0d2ffa3ce_pangea_distal_medial_tibia_1..jpg
https://upload.orthobullets.com/topic/423209/images/6e1acf6b-63f3-46d9-b5b8-dde659d59b4c_pangea_distal_medial_tibia_2..jpg
  • Pangea Design
    • Variable Angle Locking Technology
      • Pangea’s variable angle locking technology uses a CoCr locking screw, which is harder than the Ti6Al4V ELI plate, allowing for the screwhead’s threads to form a definitive locking position in the plate’s locking hole by engaging the softer, Ti6Al4V ELI material.
      • This technology allows the user to aim and lock the screw into the plate within a true 30° cone of the predetermined hole trajectory. The variable angle drill guide provided with the system offers guidance with respect to the limit of the 30° cone. The locking mechanism remains functional for up to three attempts at locking screw insertion.
      • Circular holes that accommodate screws are universal; accepting non-locking screws and locking screws within a 30° cone
    • 1.6mm K-Wire Hole
    • Oblong Hole
      • To aid in plate placement
    • K-Wire Holes at Distal Aspect
    • Single Distal Variable Angle Locking Screw Hole
      • Designed for unicortical capture of the medial malleolus
  • Placement
    • This plate is placed on the medial surface of the distal tibia.
    • Position is achieved when the distal end of the plate is over the medial malleolus.
  • Fit
    • Distal Medial Tibia plate designed for fixation of varus failure injuries of the distal tibia.
    • Designed with the use of SOMA: Stryker Orthopedics Modeling and Analytics.
    • Plate contour is SOMA designed to reduce potential for malrotation of the medial malleolus fragment. 2
    • SOMA analysis shows Distal Medial Tibia plate fits closer to bone compared to Synthes VA-LCP distal medial tibia plates. 2
  • Indications
    • The Pangea Tibia Plating System is indicated for the internal fixation and stabilization of tibia bone fractures and osteotomies in normal and osteopenic bone, including:
      • Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
      • Non-unions, malunions, and deformities
      • Periprosthetic fractures
  • Approaches
    • Surgeons may use a standard medial or anteromedial approach to the distal medial tibia. Care must be taken to protect the saphenous vein and saphenous nerve.
  • Technical specifications
    • Standard plate lengths: 3-18 hole (97-307mm)
    • Thickness: Shaft 3.3mm, Distal 1.6mm
    • Left and right anatomic plate options
    • Drill bits for T15: Ø2.5mm x 135mm (542020) Ø2.5mm x 215mm (542021)
    • Drill bits for T8: Ø2.0mm x 135mm (542000(S)) Ø2.0mm x 175mm (542001(S))
  • Screws
    • Pangea Distal Medial Tibia Plate
      • T15 Screw Platform
      • Multiaxial locking
      • 3.5 (10-120mm)
      • Cortex
      • 3.5 (10-120mm)
      • Cancellous
      • Partially threaded 4.0 (10-100mm)
      • Fully threaded 4.0 (10-100mm)
    • Pangea Distal Medial Tibia Plate
      •                                                                                                                                        T8 Screw Platform
      • Multiaxial locking
      • 2.7 (8-80mm)
      • Cortex
      • 2.7 (8-80mm)
  • References
    • 1. Pangea Distal Tibia Plating Operative Technique. PGA-ST-5, 03-2023
    • 2. Internal Report № D0000262573, Rev AA, Selzach, Switzerland