• BACKGROUND
    • Dual mobility (DM) cups of mobile polyethylene were introduced to prevent total hip arthroplasty (THA) dislocation, but no large series with this design to treat recurrent instability have been reported.
  • PURPOSE
    • Our retrospective investigation ascertained the efficiency of DM cups in correction of recurrent dislocation and assessed any adverse effects.
  • METHODS
    • One hundred eighty THAs with recurrent instability were revised to DM cups in 180 patients (mean age, 67.4 ± 11.7 years; range, 19 to 92 years). Thirty-one patients (17.2%) underwent at least one earlier THA revision, and 15 (10.3%) incurred non-union of the greater trochanter. Of the initial group in 2009, 145 patients had completed evaluations which included assessment of the Harris Hip Score and a radiographic assessment at a mean follow-up of 7.7 ± 2.2 years (range, 4 to 14 years). The rate of survival was calculated considering any reason for revision as failure.
  • RESULTS
    • At follow-up, Harris hip score was 83.9 ± 16.1 (range, 21 to 100). Dislocation of the large articulation occurred in seven hips (4.8%), and only two recurred (1.4%) (one requiring additional revision). In addition, two intra-prosthetic dislocations of the small articulation (1.4%) were observed and needed revision surgery. The large number of earlier surgeries and non-union of the greater trochanter were related to recurrent instability. Two cups (1.4%) showed signs of definite loosening; six (4.1%) presented signs of possible loosening. Twenty-nine hips manifested femoral or acetabular osteolysis (20%), but only three were severe. Eight-year survival rate considering revision for any reason was 92.6% (95% CI, 85.5-96.4%).
  • CONCLUSIONS
    • This series indicates that DM cups are a viable option to treat recurrent THA instability. Their design provides a low risk of recurrent instability without increasing mechanical complications.